Messages
Error Messages
I
-224
Illegal parameter value
The parameter value is not suitable. This error occurs when the CALC:PAR:DEF
command is used to specify an S-parameter that does not exist in the model
(S44 in the case of a 2-port model), for example.
-282
Illegal program name
This error occurs when a nonexistent VBA program name is specified by the
PROG:SEL:NAME command.
-213
Init ignored
Because another measurement is in progress, the request for initiating a
measurement ("INIT" command) is ignored.
35
Insufficient ECal module memory
This error occurs when the embedded memory is insufficient to save the user
property in ECal module during the user definition ECal processing.
-161
Invalid block data
Block data has been expected, but the block data that appears is invalid for
some reason (see 7.7.6.2, IEEE488.2). The END message is received before the
length of block data has been filled, for example.
28
Invalid calibration method
If the type of calibration is not specified or not correct when partial overwrite is
executed with the GPIB command, this error occurs. This error does not occur
for operation with front keys.
-101
Invalid character
An invalid character exists in the program message character string.
-141
Invalid character data
An invalid character is found in the character data element, or the parameter
received is not valid.
-121
Invalid character in number
A character that is invalid for the data type subject to syntactic analysis has
been received. For example, a letter is found in a decimal numeric value or a
numeric character "9" in octal data.
500
Invalid equation expression
The equation expression used in Equation Editor is not valid.
501
Invalid equation label
The equation label used in Equation Editor is not valid.