Troubleshooting
Troubleshooting Using Diagnostics Test
Table 4-4
Diagnostic test group menu
Number
7

Required Test Equipment

Table 4-5
Table 4-5
Required Equipment
Required Test Equipment
24 inch 50 Ohm cable

To Execute the diagnostics Test

To isolate faulty board assembly in analog section, execute the diagnostics test
in accordance with the following procedure. The test procedure needs to be
performed using a mouse in addition to the front panel keys.
To perform the diagnostics test properly, the following conditions must be
met:
1. Allow the analyzer to warm up for at least 90 minutes before you
2. Perform all diagnostic tests in an ambient temperature of 23°C ± 5°C.
Do not operate front panel keys, keyboard and mouse during the
diagnostics test. Changing the instrument settings while the diagnostics
test is in progress will cause incorrect test results.
Step 1. Connect a mouse to the rear panel USB connector
Step 2. Press
Step 3. Press Load Project to select Load Project function. "Open" dialog box will be
displayed as shown in
80
Test Name
RECEIVER IF RANGING TEST
shows the equipment required for performing the diagnostic test.
execute any of the diagnostic test.
key.
Figure
Qty
Recommended Model
1
Keysight p/n 8120-8862
4-9.
Keysight E5071C ENA Network Analyzers