Post-Repair Procedures
Post-Repair Procedures
Table 7-2
Post-Repair Procedures for Analog Assemblies for Over 8.5 GHz
Options
Replaced
Assembly or
Part
Card (A51)
Reference
Oven Board
(A5)
Analog Base
Module (A6)
288
Required Adjustments
Correction Constants
(CC)
PCI-DSP
Perform the following
required adjustment.
- Write ID
- Save to Flash memory
Perform the following
required adjustment.
- OCXO
Perform the following
required adjustment.
- AUX Input
- Frequency Reference
Verification
Perform
"To Execute the diagnostics
Test" on page 80
Perform the following performance test.
- AUX Input Test
- Frequency Accuracy Test
- RF Output Level Accuracy and Flatness
Test
- RF Output Level Linearity Test
- Trace Noise Test
- Crosstalk Test
- System Dynamic Range Test
- Noise Floor Test
- Dynamic Accuracy Test
- Uncorrected System Performance Test
Perform
"To Check the External
Trigger Input" on page 99
Perform
"To Execute the diagnostics
Test" on page 80
Perform the following performance test.
- Frequency Accuracy Test
Perform
"To Execute the diagnostics
Test" on page 80
Perform the following performance test.
- AUX Input Test
- Frequency Accuracy Test
Keysight E5071C ENA Network Analyzers