Advertisement

Replacement Procedure

Receiver Module Replacement for 4-Port model (A82/A87)

Receiver Module Replacement for 4-Port model (A82/A87)

1) Up to 8.5 GHz Options (A2)

Tools Required
Removal Procedure
Refer to
Step 1. Remove the outer cover as described in
Step 2. Remove the front panel as described in
Step 3. Disconnect the two semi-rigid cable (item 1).
Step 4. Disconnect the two semi-rigid cable (item 2).
Step 5. Disconnect the four semi-rigid cables (item 3).
Step 6. Remove the four TORX T10 screws fastening the module to the bottom. (item
4)
Step 7. Slide the receiver module (item 5).
248
TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end wrench, 5/8 inch
Open-end torque wrench, 5/16 inch or 8mm (set to 1.08 N-m / 9.5 lb-in)
(for reconnecting SMA connector)
Torque driver set, 5/16 inch (set to 1.08 N-m / 9.5 lb-in) (for reconnecting
SMA connector)
Figure 6-28
for this procedure.
"Outer Cover Removal" on page
"Front Panel Removal" on page
Keysight E5071C ENA Network Analyzers
203.
204.

Hide quick links: