Keysight E5061B Installation Manual
Keysight E5061B Installation Manual

Keysight E5061B Installation Manual

Network analyzer
Hide thumbs Also See for E5061B:

Advertisement

Keysight E5061B
Network Analyzer
Installation Guide

Advertisement

Table of Contents
loading

Summary of Contents for Keysight E5061B

  • Page 1 Keysight E5061B Network Analyzer Installation Guide...
  • Page 3 Network Analyzer...
  • Page 4: Safety Notices

    Do not proceed beyond a any information contained herein. Should Keysight and the user have a CAUTION notice until the indicated separate written agreement with conditions are fully understood and warranty terms covering the material met.
  • Page 5 Caution Do not exceed the operating input power, voltage, and current level and signal type appropriate for the instrument being used, refer to your instrument's Function Reference. Electrostatic discharge (ESD) can damage the highly sensitive microcircuits in your instrument. ESD damage is most likely to occur as the test fixtures are being connected or disconnected.
  • Page 6: Safety Summary

    When you notice any of the unusual conditions listed below, immediately terminate operation and disconnect the power cable. Contact your local Keysight Technologies sales representative or authorized service company for repair of the instrument. If you continue to operate without repairing the instrument, there is a potential fire or shock hazard to the operator.
  • Page 7 Manufacturer’s Declaration Herstellerbescheinigung GERA- USCHEMISSION LpA < 70 dB am Arbeitsplatz normaler Betrieb nach DIN 45635 T. 19 Manufacturer's Declaration ACOUSTIC NOISE EMISSION LpA < 70 dB operator position normal operation per ISO 7779 Network Analyzer...
  • Page 8: Regulatory Compliance Information

    This product complies with the essential requirements of the following applicable European Directives, and carries the CE marking accordingly: • The Low Voltage Directive 2006/95/EC • The EMC Directive 2004/108/EEC To obtain Declaration of Conformity, please contact your local Keysight Technologies sales office, agent or distributor. Network Analyzer...
  • Page 9: Safety Notice Supplement

    • This equipment is a POLLUTION DEGREE 2, INDOOR USE product. • This equipment is tested in stand-alone condition and in combination with the accessories supplied by Keysight Technologies against the requirement of the standards described in the Declaration of Conformity. If it is used...
  • Page 10: General Safety Precautions

    N O T E POLLUTION DEGREE 2 in IEC61010-1. The E5061B is an INDOOR USE product. The LEDs in the E5061B are Class 1 in accordance with IEC60825-1, N O T E CLASS 1 LED PRODUCT. • Ground the Instrument To avoid electric shock, the instrument chassis and cabinet must be grounded with the supplied power cable’s...
  • Page 11 To avoid the danger of introducing additional hazards, do not install substitute parts or perform unauthorized modifications to the instrument. Return the instrument to a Keysight Technologies Sales and Service Office for service and repair to ensure that safety features are maintained in operational condition.
  • Page 12: Safety Symbols

    Safety Symbols General definitions of safety symbols used on the instrument or in manuals are listed below. Instruction Manual symbol: the product is marked with this symbol when it is necessary for the user to refer to the instrument manual. Alternating current.
  • Page 13 Keysight shall not be liable for errors or for incidental or consequential damages in connection with the furnishing, use, or performance of this document or any information contained herein.
  • Page 14 The installation guide (this manual) provides start up setup information when you use the E5061B for the first time and troubleshooting information when the Windows cannot be boot up. See this manual first when you use the E5061B for the first time. Online Help...
  • Page 15 Help key in the E5061B or by pressing F1 in a keyboard attached to the E5061B or by clicking the help button in a dialog box. With context sensitive help, users can receive information quickly about the area the user is working in the firmware of the E5061B.
  • Page 16 The following shows the contents of this manual. Chapter 1, "Installation" This chapter provides information about how to set up the Keysight E5061B. Chapter 2, "Troubleshooting" This chapter describes the troubleshooting during start up and the procedure of the operating system (OS) recovery when the Windows XP OS has been damaged.
  • Page 17: Table Of Contents

    Verification of the Power Supply Verification and Connection of Power Cable Blown Fuses Starting the E5061B Turning the Power ON and OFF Disconnection from Supply Source Initial Registration of E5061B Calibration of the Touch Screen Troubleshooting Troubleshooting during Startup System Recovery Network Analyzer...
  • Page 18 Notes on executing the factory recovery function Procedure to execute the factory recovery Index Regional Sales and Support Offices Network Analyzer...
  • Page 19: Installation

    Installing Front Handles/Rack Mounting Flanges 23 Connecting the Accessories 27 Power Supply 28 Blown Fuses 29 Starting the E5061B 30 Initial Registration of E5061B 33 Calibration of the Touch Screen 46 This chapter provides information about how to set up the Keysight E5061B.
  • Page 20: Checking The Shipment

    Step 4. After checking, if one of the following applies, contact your nearest Keysight Technologies sales and service office. The packing box or shock-absorbing material used to package the analyzer has been damaged or the shock-absorbing material has traces where extreme force has been applied.
  • Page 21 Installation inspection by the transport company, save the packing box, shock-absorbing material, and packaged items as you received them. Network Analyzer...
  • Page 22: Environmental Requirements

    For more information on environmental temperature to satisfy the specifications and measurement accuracy of the analyzer, see the specification in the E5061B Online Help. When the environmental temperature around the analyzer is...
  • Page 23: Protection Against Electrostatic Discharge (Esd)

    ±42V Peak Max Output e5061b060 Figure 1 Ventilation space at the installation Place the E5061B in a proper position as show in Figure N O T E Protection Against Electrostatic Discharge (ESD) Set up a static-safe work-station to protect the electronic...
  • Page 24: Ensuring Adequate Free Space Around Analyzer For Immediate Disconnection Of Power Cable In Case Of Emergency

    “Disconnection from Supply Source" on page the power supply is disconnected by removing the power cable’s connector plug from either the AC outlet or the E5061B unit. When installing the E5061B, ensure that there is sufficient free space around the unit to permit quick disconnection of the plug (from AC outlet or E5061B unit) in case of emergency.
  • Page 25: Installing Front Handles/Rack Mounting Flanges

    Installation Installing Front Handles/Rack Mounting Flanges The E5061B can be installed on a workbench or in a rack. This section describes how to install the front handles (Option 1CN) used for moving or transporting the instrument, and how to install the analyzer in an equipment rack as part of a measurement system (Option 1CM: without the handles, Option 1CP: with the handles).
  • Page 26: How To Install The Handle Kit (Option 1Cn)

    Step 1. Remove the adhesive-backed trim strip (1) from each side of the outer frame of the E5061B front panel. Step 2. Use the screws provided screws to mount the front handles (2) on each side of the E5061B front panel frame.
  • Page 27: How To Install The Rack-Mount Kit (Option 1Cm)

    Step 1. Remove the adhesive-backed trim strip (1) from each side of the outer frame of the E5061B front panel. Step 2. Use the screws provided to mount a rack-mounting flange (4) on each side of the E5061B front panel frame.
  • Page 28 Installation Step 3. Remove the four bottom feet of the E5061B (lift the bar marked TAB on the inner side of the foot and slide the foot toward the bar). Step 4. Mount the E5061B on the rack. Network Analyzer...
  • Page 29: Connecting The Accessories

    Figure 4 Connecting the BNC adaptor Connecting Mouse and Keyboard The E5061B allows you to connect Mouse and/or keyboard through USB. USB mouse and keyboard can be connected with the USB ports on the front or rear panels. Initial registration of the E5061B requires the mouse and keyboard before turning on the power.
  • Page 30: Power Supply

    Maximum power 300 VA consumption * 1. Switched automatically by the E5061B in conformity to the voltage used. Verification and Connection of Power Cable The three-wire power cable attached to the E5061B has one wire serving as a ground. Using this power cable allows the E5061B to be grounded, thereby protecting you against electrical shock from the power outlet.
  • Page 31: Blown Fuses

    If the fuse appears to have blown during operation, this instrument may be subject to failure and must be repaired. For any assistance, contact Keysight Technologies Customer contact center listed at the end of this guide. The product uses the following fuse types: UL/CSA Type, Slow-Blo, 10 A-250 Vac.
  • Page 32: Starting The E5061B

    Installation Starting the E5061B This section describes how to turn on/off the E5061B power and how to cut off the power supply in an emergency. Turning the Power ON and OFF The standby switch can turn/off the E5061B. The color on...
  • Page 33: Disconnection From Supply Source

    Line switch (Always ON) and power cable receptacle Disconnection from Supply Source The power supply of the E5061B is cut off by disconnecting the plug of the power cable (on either AC outlet side or E5061B side). When it is necessary to disconnect the power supply in order to avoid shock hazards, etc., pull out the...
  • Page 34 Installation To allow this operation to be performed smoothly, be sure to follow the N O T E guidelines in “Ensuring Adequate Free Space around Analyzer for Immediate Disconnection of Power Cable in Case of Emergency" on page When turning the power OFF under normal circumstances, always follow the methods described in “Turning the Power OFF"...
  • Page 35: Initial Registration Of E5061B

    N O T E - Windows Vista Business - Windows XP Pro for Embedded Systems The license is shown at the Windows Lincese label on the E5061B rear panel. You have different initial registration procedure depending on the windows license.
  • Page 36 Installation Windows XP Pro for Embedded Systems (HDD Rev KY22x and above) Step 1. Turn on the E5061B. Do not touch any key on the front panel during boot up. Step 2. The screen as shown in the Figure 6 appears.
  • Page 37 Installation Figure 7 System start up screen in Windows XP Pro for Embedded Systems (2/2) Step 4. The windows start up screen as shown in Figure 8 appears. Network Analyzer...
  • Page 38 Installation Figure 8 Windows XP Pro for Embedded Systems start up screen Step 5. In the Keysight Tech. Software End User License Agreement dialog box, select the Agree button. Figure 9 shows the license agreement dialog box in Windows XP Pro for Embedded Systems (for HDD Rev KY22x and above).
  • Page 39 Figure 9 Windows XP Pro for Embedded Systems license agreement dialog box Step 6. After a while, the E5061B restarts automatically. Windows XP Pro for Embedded Systems (HDD Rev KY21x and below) Step 1-4. Follow the steps of Windows XP Pro for Embedded Systems (HDD Rev KY22x and above).
  • Page 40 Installation Figure 10 Windows XP Pro for Embedded Systems license agreement dialog box Network Analyzer...
  • Page 41 Windows. Then, click the Next> button (Figure 11). Figure 11 Date and Time Setting dialog box Step 7. After a while, the E5061B restarts automatically. Windows Vista Business Step 1. Follow the steps of Windows XP Pro for Embedded Systems (HDD Rev KY22x and above).
  • Page 42 Installation Step 2. The screen as shown in the Figure 12 appears. Select Microsoft Windows XP Professional and press [Enter]. Figure 12 System start up screen in Windows Vista Business (1/2) After several seconds, the next screen appears automatically even if you N O T E do not press any key.
  • Page 43 Installation Figure 13 System start up screen in Windows Vista Business (2/2) Step 4. The Windows XP Professional Setup wizard appears. Click the Next> button (Figure 14). Network Analyzer...
  • Page 44 Installation Figure 14 Windows XP Professional Setup wizard Network Analyzer...
  • Page 45 Installation Step 5. In the License agreement dialog box, select the I accept this agreement box and click the Next > button. Figure 15 shows the license agreement dialog box in Windows Vista Business. Figure 15 Windows Vista Business license agreement dialog box Network Analyzer...
  • Page 46 Figure 16 Date and Time Setting dialog box Step 7. After a while, the E5061B restarts automatically. The Keysight Recovery Image Setup dialog box as shown in Figure 17 appears. Click the OK button. Figure 17 Keysight Recovery Image Setup...
  • Page 47 Installation Step 8. Once the recovery image setup is completed, the E5061B restarts again. Network Analyzer...
  • Page 48: Calibration Of The Touch Screen

    Installation Calibration of the Touch Screen When E5061B measurement screen appears, you have to calibrate the touch screen. Follow the procedure described below to calibrate the touch screen. Step 1. Press Step 2. Press Service Menu. Step 3. Press Test Menu.
  • Page 49 Installation If you do not start to perform the touch screen calibration within several N O T E seconds, the previous measurement screen re-appears automatically. Network Analyzer...
  • Page 50 Installation Network Analyzer...
  • Page 51: Troubleshooting

    Keysight E5061B Network Analyzer Installation Guide Troubleshooting Troubleshooting during Startup 50 System Recovery 51 This chapter describes the troubleshooting process during start up and the procedure of the operating system (OS) recovery when the Windows XP OS has been damaged.
  • Page 52: Troubleshooting During Startup

    Troubleshooting Troubleshooting during Startup When you encounter problems during start up, see Table System recovery saves most of problems. Table 3 Troubleshooting during startup Symptom Solution Turning on (|) the standby switch does Confirm that the power cable is not start up the system. properly plugged in.
  • Page 53: System Recovery

    D drive) are not affected. However, Keysight recommends backing them up before executing system recovery for precautionary purposes. For more information on backup, refer to “Backing Up the Data” as described in E5061B Online Help. Procedure to execute the factory recovery This section describes how to return the contents of the C drive to the factory state.
  • Page 54 Step 3. Connect the keyboard to the E5061B. Step 4. Press the standby switch of the E5061B to turn it Windows Vista Business Step 5. When the screen as shown in the figure below appears, select Keysight Recovery System and press [Enter].
  • Page 55 Troubleshooting Step 6. Type 2 to select Recover Factory Backup Image option and press [Enter]. Figure 20 System recovery selection screen Step 7. A confirmation message appears. Type 1 and press [Enter] to continue the process. Figure 21 System recovery confirmation screen (1/2) Network Analyzer...
  • Page 56 Once the recovery process is completed, completion message prompts. Press any key to continue. The E5061B restarts automatically. Figure 23 Recovery Option confirmation dialog box After several seconds, the E5061B restarts automatically even if you do N O T E not press any key. Network Analyzer...
  • Page 57 Windows XP Pro for Embedded Systems Step 5. When the screen as shown in the figure below appears, select Keysight Recovery System and press [Enter]. Figure 24 Keysight Recovery System selection screen After several seconds, the next screen appears automatically even if you N O T E do not press any key.
  • Page 58 Troubleshooting Step 7. A confirmation message appears. Click OK to continue the process. Figure 26 System recovery confirmation screen Step 8. The progress of the system configuration is displayed on the screen. The recovery takes a few minutes depending on the amount of data. Figure 27 System configuration progress screen Network Analyzer...
  • Page 59 “Calibration of the Touch Screen" on page Never turn off the power during the system recovery because doing C A U T I O N so may cause serious damage to the E5061B. Network Analyzer...
  • Page 60 Troubleshooting Network Analyzer...
  • Page 61 23 System Recovery initial registration, 27, 33 how to execute factory recovery, 51 inspection how to execute user recovery function, 55, checking the shipment, 18 types of system recoveries, 51 Keysight Technologies sales and service office Network Analyzer...
  • Page 62 Index Network Analyzer...
  • Page 64 This information is subject to change without notice. © Keysight Technologies 2010, 2011, 2013, 2014 Edition 4, November 2014 *E5061-90017* E5061-90017 www.keysight.com...

Table of Contents