Replacement Procedure

A4/A5 Local Synthesize Module Replacement

A4/A5 Local Synthesize Module Replacement

Tools Required

TORX screwdriver, T10, T15, and T20
Flat edge screwdriver
Open-end torque wrench, 5/16 inch (set to 1.02 N-m / 9.0 lb-in) (for reconnecting SMA
connector)

Removal Procedure

Refer to
Figure 6-20
Step 1. Remove the outer cover as described in
Step 2. Remove the front panel as described in
Step 3. Disconnect the semi-rigid cables (item 1).
Step 4. Remove each two TORX T10 screws (item 2) from the local synthesize module.
Step 5. Slide out the local synthesize module (item 3).
Step 6. Release the clamped cables from the clamp (item 4).
Step 7. Disconnect the RF cable (item 5) from the local synthesize module.