Replacement Procedure; Replacing The Probe Tip - Keysight 85024A User's And Service Manual

High frequency probe
Table of Contents

Advertisement

Replacement Procedure

Replacing the probe tip

Tools required:
– 3/32 inch probe-tip nut driver (supplied with the probe)
– 10 mm open-end wrench
Use static precautions when performing the following procedures. The
CAUTION
amplifier microcircuit is very static sensitive and exposed during this
procedure.
Refer to
1 Remove the nut with a 10 mm open-end wrench.
2 Remove the nose assembly, leaving the microcircuit attached to the body of
Failure to remove the nose assembly prior to replacing the probe tip will
CAUTION
result in damage to the conductive elastomer on the new probe tip
assembly.
3 Unscrew the damaged tip with the nut driver and discard the tip.
4 Screw in the new tip and lightly tighten it with a 3/32 inch nut driver.
5 Install the nose assembly and nut. Tighten the nut with a 10 mm open-end
The tip on the 10:1 divider can be replaced without disassembling the divider.
NOTE
Use the 3/32 inch probe tip driver.
Keysight 85024A User's and Service Guide
Figure
7-6.
the probe.
(Over-tightening the tip can damage the nose assembly.)
wrench.
Service
7
65