Optional glass capillary sprayer

The glass capillary sprayer is designed for use with metal-coated borosilicate
glass capillaries. They allow extremely low flow rates (less than 100 nL/min).
The glass capillaries are used for one sample only and must then be discarded.
To use the glass capillary sprayer, complete the following procedures:
Plumb the backpressure gas.
Install the glass capillary sprayer.
Fit and load the glass capillary.
Optimize the sprayer.
Plumbing the backpressure gas
The electronic pressure controller can control pressure only when gas is
flowing. However, the glass capillary requires a backpressure without a flow.
To resolve this situation, the backpressure gas is plumbed from the source to
the glass capillary sprayer through an in-line tee-piece. The tee-piece permits
the gas flow to bleed to atmosphere while the selected pressure is maintained
at the back of the capillary.
Required materials
Chemical-resistant, powder-free gloves
Glass capillary option spares kit, which includes these items:
PTFE tubing, 1/16-inch OD, 0.006-inch ID, 530 mm
PTFE tubing, 1/16-inch OD, 0.023-inch ID, 120 mm
Tee piece and nuts, 1/16-inch "finger-tight", 0.5-mm bore
SealTight™ nut and ferrule
Sharp knife or PEEK tubing cutter
To plumb the backpressure gas:
1.
Use the sharp knife or PEEK tubing cutter to cut a 30-mm length from
the 0.006-inch ID PTFE tubing to leave a 500-mm length, and retain
both pieces.
Optional glass capillary sprayer
4-9