8.
Slide open the instrument's source interface door (see the figure on
page
1-5).
9.
Connect the probe cable to the instrument's PROBE connection.
10. Connect the high voltage cable to the instrument's HV connection.
High voltage cable
11. Slide closed the instrument's source interface door.
Fitting a borosilicate glass capillary (nanovial)
Required materials
•
Chemical-resistant, powder-free gloves
•
Needle-nose pliers
•
Borosilicate glass capillary
The NanoFlow stage contains a high voltage interlock so
Caution:
that unless the sprayer is pushed fully forward in the source, the
capillary voltage (the voltage applied to the sprayer assembly) and
the sampling cone voltage are disabled.
NanoFlow ESI source
3-21