Replacing the ESI probe tip and gasket
Replace the ESI probe tip if a blockage occurs in the internal metal sheathing
through which the stainless steel capillary passes or if the probe tip is
damaged.
Removing the ESI probe tip and gasket
Required materials
•
Chemical-resistant, powder-free gloves
•
7-mm wrench
•
10-mm wrench
To remove the ESI probe tip and gasket:
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
great care while performing this procedure.
Warning:
the probe with care.
The probe and source components can be contaminated
The probe and source can be hot. To avoid burn injuries, take
The ESI probe tip is sharp. To avoid puncture wounds, handle
Replacing the ESI probe tip and gasket
5-63