Replacing the ESI probe tip and gasket

Replace the ESI probe tip if a blockage occurs in the internal metal sheathing
through which the stainless steel capillary passes or if the probe tip is
damaged.

Removing the ESI probe tip and gasket

Required materials
Chemical-resistant, powder-free gloves
7-mm wrench
10-mm wrench
To remove the ESI probe tip and gasket:
Warning:
with biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.
Warning:
great care while performing this procedure.
Warning:
the probe with care.
1.
Remove the ESI probe from the source (see
The probe and source components can be contaminated
The probe and source can be hot. To avoid burn injuries, take
The ESI probe tip is sharp. To avoid puncture wounds, handle
page
3-4).
Replacing the ESI probe tip and gasket
4-65