4 Maintenance Procedures

Replacing the ESI probe tip and gasket

Replace the ESI probe tip if a blockage occurs in the internal metal sheathing
through which the stainless steel capillary passes or if the probe tip is
damaged.

Removing the ESI probe tip and gasket

Required materials
Chemical-resistant, powder-free gloves
7-mm wrench
10-mm wrench
To remove the ESI probe tip and gasket:
1.
Remove the ESI probe from the source (see
page
56).
146
February 11, 2013, 71500130002 Rev. D
To avoid personal contamination with biohazards
Warning:
or compounds that are toxic or corrosive, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the probe and source components.
To avoid burn injuries, take great care while
Warning:
performing this procedure; the probe and source can be hot.
To avoid puncture wounds, handle the probe with
Warning:
care; the ESI probe tip is sharp.
"Removing the ESI probe" on