4 Maintenance Procedures
Replacing the ESI probe tip and gasket
Replace the ESI probe tip if a blockage occurs in the internal metal sheathing
through which the stainless steel capillary passes or if the probe tip is
damaged.
Removing the ESI probe tip and gasket
Required materials
•
Chemical-resistant, powder-free gloves
•
7-mm wrench
•
10-mm wrench
To remove the ESI probe tip and gasket:
1.
Remove the ESI probe from the source (see
page
56).
146
February 11, 2013, 71500130002 Rev. D
To avoid personal contamination with biohazards
Warning:
or compounds that are toxic or corrosive, and to avoid
spreading contamination to uncontaminated surfaces, wear
clean, chemical-resistant, powder-free gloves when working
with the probe and source components.
To avoid burn injuries, take great care while
Warning:
performing this procedure; the probe and source can be hot.
To avoid puncture wounds, handle the probe with
Warning:
care; the ESI probe tip is sharp.
"Removing the ESI probe" on