Replacing The Esi Probe Tip And Gasket; Removing The Esi Probe Tip And Gasket - Waters Xevo TQD IVD Overview And Maintenance Manual

Table of Contents

Advertisement

6.
Fit the source enclosure onto the instrument (see

4.22 Replacing the ESI probe tip and gasket

Replace the ESI probe tip if a blockage occurs in the internal metal sheathing through which the
stainless steel capillary passes or if the probe tip is damaged.
4.22.1

Removing the ESI probe tip and gasket

4.22.1.1
Required materials
Chemical-resistant, powder-free gloves
7-mm wrench
10-mm wrench
Warning:
To avoid personal contamination with biohazards or toxic materials, and to avoid
spreading contamination to uncontaminated surfaces, wear clean, chemical-resistant, powder-free
gloves when working with the probe and source components.
Warning:
and source can be hot.
Warning:
To remove the ESI probe tip and gasket:
1.
Remove the ESI probe from the source (see
To avoid burn injuries, take great care while performing this procedure; the probe
To avoid puncture wounds, handle the probe with care; the ESI probe tip is sharp.
September 7, 2016, 715003560IVD Rev. C
page
46).
page
39).
Page 93