Replacing the ESI probe tip and gasket
Replace the ESI probe tip if a blockage occurs in the internal metal sheathing
through which the stainless steel capillary passes or if the probe tip is
damaged.
Removing the ESI probe tip and gasket
Required materials
•
Chemical-resistant, powder-free gloves
•
7-mm wrench
•
10-mm wrench
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the probe and source
components.
Warning:
this procedure; the probe and source can be hot.
Warning:
the ESI probe tip is sharp.
To remove the ESI probe tip and gasket:
1.
Remove the ESI probe from the source (see
To avoid personal contamination with biohazards or
To avoid burn injuries, take great care while performing
To avoid puncture wounds, handle the probe with care;
March 14, 2014, 715004387 Rev. B
Replacing the ESI probe tip and gasket
page
61).
147