To adjust the grip of the sample cup holder
Use tweezers to squeeze the sample cup holder prongs together, or use a
jeweler's screwdriver to gently pry the prongs apart.

Removing the sample cup from the probe tip

Warning:
materials. Always wear nitrile gloves while handling the components.
To remove the sample cup from the probe tip
Use tweezers to remove the sample cup from the sample cup holder
If it is difficult to remove the sample cup from the sample cup holder, use
Tip:
a jeweler's screwdriver to gently pry the prongs apart.

Inserting the solids sample introduction probe in the ion source

Warning:
toxic materials. Wear nitrile gloves at all times while handling the
components.
To insert the solids sample introduction probe in the ion source
1.
Ensure that the probe handle is fully retracted.
The probe components can be contaminated with toxic
The source and probe components can be contaminated with
The solids sample introduction probe
6-9