Disassembling The Sampling Cone Assembly - Waters Xevo TQD IVD Overview And Maintenance Manual

Table of Contents

Advertisement

2.
Grasp the cone gas nozzle handle, and use it to rotate the sampling cone assembly 90
degrees, moving the handle from the vertical to the horizontal position.
Notice:
isolation valve at any time while the sampling cone assembly is removed from the ion
block assembly.
3.
Slide the sampling cone assembly out of the ion block assembly.
Ion block assembly
4.18.2

Disassembling the sampling cone assembly

4.18.2.1
Required material
Chemical-resistant, powder-free gloves
To avoid damaging the instrument by sudden venting, do not open the
September 7, 2016, 715003560IVD Rev. C
Page 65
Sampling cone
assembly
Cone gas nozzle
handle
TP03131