3. Slide the sampling cone assembly out of the ion block assembly.
Figure 5–12: Removing the sampling cone assembly
Ion block assembly

5.13.2 Disassembling the sampling cone assembly

Required materials
• Chemical-resistant, powder-free gloves
• Combined 2.5-mm hex wrench and cone extraction tool
To disassemble the sampling cone assembly:
1. Retrieve the combined 2.5-mm hex wrench and cone extraction tool from its storage
location behind the source enclosure.
Figure 5–13: Cone extraction tool location
Combined 2.5-mm hex wrench and cone extraction tool
2. Slide the collar to the end of the tool.
Notice:
To avoid damage, do not open the source isolation valve before fitting
the sampling cone assembly to the ion block assembly.
March 17, 2021, 715005046 Ver. 03