Disassembling the sampling cone assembly

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
To disassemble the sampling cone assembly:
1.
On the combined 2.5-mm Allen wrench and cone extraction tool, slide
the collar to the end of the tool.
Collar
2.
Insert the collar in the sample cone.
4-40
Maintenance Procedures
The sampling cone assembly can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.