Disassembling The Sampling Cone Assembly - Waters Xevo TQ MS Operator And Maintenance Manual

Table of Contents

Advertisement

Disassembling the sampling cone assembly

Required material
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
To disassemble the sampling cone assembly
1.
On the combined 2.5-mm Allen wrench and cone extraction tool, slide
the collar to the end of the tool.
Collar
2.
Insert the collar in the sample cone.
4-28
Maintenance Procedures
The sampling cone assembly can be contaminated with
Warning:
biohazardous and/or toxic materials. Always wear
chemical-resistant, powder-free gloves while performing this
procedure.

Advertisement

Table of Contents
loading

Table of Contents