Medical Gas Central Alarm - Medipoint 125 Digital Alarm
Figure 44 - Data logging Screen, Filtered by Gas
service.
Special Characters/Icons:
Icon
Description
Previous Icon.
Takes you to the previous logs screen.
Next Icon.
Takes you to the next logs screen.
Down Icon.
Jumps down 50 records on the current logs screen
Up Icon.
Jumps up 50 page on the current logs screen
Scroll Bar.
Press, hold and drag the scroll bar navigate between
the selected 50 records.
Exit Icon.
Takes you back to the main screen.
Press the exit
icon to return back to the main screen.
Note...
• The system will return back to the main screen
automatically after five minute if the alarm is not
manually exited from the information screen.
• If an alarm status becomes active while a technician
is viewing the logs the audible alarm will activate
to make the user aware of a new condition.
4233500107.03
6.

Maintenance Procedures.

A competent person who is conversant with the
maintenance of medical gas pipeline equipment and
any special national safety requirements which may
apply must carry out all maintenance. Preventative
maintenance contracts are available from BeaconMedæs
for installations within the U.K. Overseas distributors will
be able to supply similar contracts in other areas.
CAUTION! It is essential that only genuine
BeaconMedæs spare parts are fitted during maintenance.
WARNING:
equivalent for overseas) before commencing with
any work on a medical gas installation.
6.1
Replacement of alarm panel
components.
WARNING! Ensure the power supply is
safely isolated before accessing the alarm panel for
maintenance or repairs.
WARNING! The isolation and connection to
the electrical power supply should only be carried
out by a suitably qualified electrician.
NOTE - During electrical isolation any connected repeater
alarm panels will show a system fault.
6.1.1
Isolate the power supply to the alarm panel.
6.1.2
Turn the security fastener with the supplied
key tool access screw counterclockwise until the hinged
door can be opened.
6.1.3
See figure 45a & 45b for alarm panel assembly
details for replacement of the alarm component.
CAUTION!
susceptible to damage by static electricity and must
remain enclosed in their anti-static packaging until
immediately required for use. To prevent damage to
printed circuit boards, handle with care and do not
over torque retaining nuts.
6.1.4
Close the alarm panel door and fasten the
security screw by turning clockwise.
6.1.5
Return power to the alarm panel and complete
the commissioning section 4.
32
Obtain
a
work
permit
Printed