Wiring

Power Supply

Alarm panels require 100 to 250 VAC
50/60 Hz. power. Refer to Wiring
Schematic 1 (Page. 46).
WARNING:
RISK OF ELECTRIC SHOCK
Disconnect power at the circuit breaker before
removing power supply shield.
For NFPA 99 compliance, alarm panel
must be connected to life safety branch of
the emergency electrical system.
1. Remove four nuts (area alarms) or
screws (master or combo alarms) from
plastic power supply shield.
2. Remove plastic shield from power
supply.
3. Connect incoming line, neutral and
ground wires to the terminal block.
5. Reinstall plastic power supply shield
while making sure all high voltage wires
are contained within plastic shield.
6. Secure plastic shield with four nuts
(area alarms) or screws (master or
combo alarms).
30