5 Function check
Result
5.4 Checking the reusable breathing
Requirement
Result
52
EN
MEDUVENT Standard
17. Select Finish.
18. Switch off the device.
19. Disconnect the test lung from the breathing circuit.
The function check is complete.
circuit
The patient valve of the reusable breathing circuit has been
removed
(see "8.3.1 Disassembling the reusable breathing
circuit", page
87).
1. Check all parts of the patient valve for external damage.
2. If necessary: Replace damaged parts.
3. Check the PEEP control diaphragm and check valve diaphragm:
4. If the diaphragm is torn, corrugated, distorted or sticky:
Replace the diaphragm.
5. Assemble the reusable breathing circuit
the reusable breathing circuit", page
6. Check the system for leaks with a new function check.
The reusable breathing circuit has been checked.
(see "8.3.2 Assembling
90).