5 Function check

5.1 Intervals

5.2 Checking the portable unit visually

Result
82
EN
LIFE-BASE
Part
Portable unit including all attached
devices
1. Check the portable unit for external damage.
2. If available: Check the charging interface and
connection line for:
Damage
Wear
Exposed wires
Bent connection lines
3. If available: Supply the portable unit with voltage via
the charging interface.
When doing so, note:
The LED for the external power supply must light up.
The battery status indicator must light up or flash.
4. If parts are damaged: Either have the portable unit repaired
by WEINMANN Emergency or replace it.
The portable unit has been checked visually.
Interval
Before each use
After each assembly
After each hygienic reprocessing
At least every 6 months