5. Test the device

5.1 General

Important
A test is required after every service or repair.
Please enter the following figures in your service
record sheet (see Page 53):
operating hours, humidifier
For SOMNOsmart 2:
level, Softstart time and Softstart initial pressure,
and the patient's pressure limits and pressure rise
rate as shown in the patient pass.
operating hours, humidifier level,
For SOMNOset:
Softstart time and Softstart initial pressure, and the
patient's pressure limits and pressure rise rate as
shown in the patient pass, and also the mode.
+
operating hours, humidifier
For SOMNOsoft
:
level, Softstart time and Softstart initial pressure,
CPAP pressure and mode.

5.2 Performing the check

5.2.1 Checking the power cord

1. Check the power cord
Make sure that
– the insulation is sound,
– the cable is undamaged,
– there are no loose connections.
2. If necessary, change the power cord

5.2.2 Checking the housing

Check the general condition of the housing.
– If the housing is damaged or faulty, replace
the relevant side of the device (see „7.15
Replacing lower part of housing" on
page 41 or „7.16 Replacing upper part of
housing" on page 42).
.
3
.
3
If the test reveals faults or deviations from the
specified values, the device must not be used
again until the faults are rectified.
To find out what may be causing the faults and
how to remedy the malfunction, see Chapter „6.
Troubleshooting" on page 20.