18. Switch off the device.
19. Disconnect the test lung from the breathing circuit.
Result
The function check is complete.
5.4 Checking reusable breathing circuit
Requirements
The patient valve of the reusable breathing circuit has been
disassembled
circuit", page
1. Check all parts of the patient valve for external damage.
2. If necessary: Replace damaged parts.
3. Check the PEEP control diaphragm and check valve diaphragm.
4. Assemble the reusable breathing circuit
5. Check the system for leaks with a new function check.
Result
The reusable breathing circuit has been tested.
(see "7.3.1 Disassembling the reusable breathing
74).
If a diaphragm is torn, corrugated, distorted or sticky: Replace
the diaphragm.
"7.3.2 Assembling the reusable breathing circuit", page
(see
MEDUVENT Standard
5 Function check
77).
EN
53