APPENDIX A-3:
Table A-2:
Name
WSYSRES
SYSTEM RESET
WDATAINIT
DATA INITIALIZED
WCONFIGINIT
CONFIG INITIALIZED
WSAMPFLOW
SAMPLE FLOW WARN
WOZONEFLOW
OZONE FLOW
WARNING
WOZONEGEN
OZONE GEN OFF
WRCELLPRESS
RCELL PRESS WARN
WBOXTEMP
BOX TEMP WARNING
WRCELLTEMP
RCELL TEMP WARNING
4
WMANIFOLDTEMP
MANIFOLD TEMP
WARN
WIZSTEMP
IZS TEMP WARNING
WCONVTEMP
CONV TEMP WARNING
WPMTTEMP
PMT TEMP WARNING
WAUTOZERO
AZERO WRN XXX.X MV
WHVPS
HVPS WARNING
WDYNZERO
CANNOT DYN ZERO
WDYNSPAN
CANNOT DYN SPAN
WREARBOARD
REAR BOARD NOT DET
WRELAYBOARD
RELAY BOARD WARN
WFRONTPANEL
FRONT PANEL WARN
WANALOGCAL
ANALOG CAL
WARNING
1
Optional equipment only
044100110 Rev A

Warnings and Test Measurements, Revision C.8

M200E Warning Messages, Revision C.8
Message Text
Instrument was power-cycled or the CPU was reset.
Data storage was erased.
Configuration storage was reset to factory configuration or
erased.
Sample flow outside of warning limits specified by SFLOW_SET
variable.
Ozone flow outside of warning limits specified by OFLOW_SET
variable.
Ozone generator is off. This is the only warning message that
automatically clears itself. It clears itself when the ozone
generator is turned on.
Reaction cell pressure outside of warning limits specified by
RCELL_PRESS_SET variable.
Chassis temperature outside of warning limits specified by
BOX_SET variable.
Reaction cell temperature outside of warning limits specified by
RCELL_SET variable.
Bypass or dilution manifold temperature outside of warning limits
specified by MANIFOLD_SET variable.
IZS temperature outside of warning limits specified by IZS_SET
variable.
Converter temperature outside of warning limits specified by
CONV_SET variable.
PMT temperature outside of warning limits specified by PMT_SET
variable.
Auto-zero reading above limit specified by AZERO_LIMIT
variable. Value shown in message indicates auto-zero reading at
time warning was displayed.
High voltage power supply output outside of warning limits
specified by HVPS_SET variable.
Contact closure zero calibration failed while DYN_ZERO was set
to ON.
Contact closure span calibration failed while DYN_SPAN was set
to ON.
Rear board was not detected during power up.
Firmware is unable to communicate with the relay board.
Firmware is unable to communicate with the front panel.
The A/D or at least one D/A channel has not been calibrated.
APPENDIX A
Description
241