Troubleshooting & Repair
Attach the outlet port of a flow meter to the sample inlet port on the rear panel.
Ensure that the inlet to the flow meter is at atmospheric pressure.
The sample flow measured with the external flow meter should be 500 cm³/min ±
10%. If a combined sample/ozone air Perma Pure dryer is installed (optional equip-
ment), the flow will be 640 cm³/min ± 10% (500 cm³/min for the sample and 80
cm³/min for the ozone generator supply air and 60 cm³/min for the purge flow).
Low flows indicate blockage somewhere in the pneumatic pathway.

11.5.4. AC Power Configuration

The M200E can be easily configured for two main power regimes, 100-120 V and 220-240 V
at either 50 or 60 Hz. The analyzer is correctly configured for the AC power voltage in use if
the voltage selector plug is configured properly. See Figure 3-1 and Figure 11-4 for the
location of this plug. The plug for 100-120 V should have white jumper wires, the plug for
220-240 V should have blue jumper wires. Note, that an analyzer configured for 230 V will
still turn on at 115 V (and vice versa), but the heaters may burn out or not heat up fast
enough. Internally, several LEDs should turn on as soon as the power is supplied. If an
incorrect power configuration is suspected, check for the correct voltage and frequency at
the line input on the rear panel.
Note that the analyzer will be severely damaged if 220-240 V is supplied to it when
configured for 100-120 V. Never bypass the power switch or circuit breaker.

11.5.5. DC Power Supply

If you have determined that the analyzer's AC main power is working, but the unit is still
not operating properly, there may be a problem with one of the instrument's switching
power supplies, which convert AC power to 5 and ±15 V (PS1) as well as +12 V DC power
(PS2). The supplies can either have no DC output at all or a noisy output (fluctuating).
To assist tracing DC Power Supply problems, the wiring used to connect the various printed
circuit assemblies and DC powered components and the associated test points on the relay
board follow a standard color-coding scheme as defined in Table 11-3.

Table 11-3: DC Power Test Point and Wiring Color Code

Name
DGND
+5V
AGND
+15V
-15V
+12R
+12V
A voltmeter should be used to verify that the DC voltages are correct as listed in
Table 11-5. An oscilloscope, in AC mode and with band limiting turned on, can be used to
evaluate if the supplies are excessively noisy (>100 mV peak-to-peak).
206
Test Point#
Color
1
Black
2
Red
3
Green
4
Blue
5
Yellow
6
Purple
7
Orange
Model 200E Instruction Manual
Definition
Digital ground
Analog ground
12 V return (ground) line
044100102 Rev A