Trigger:glitch Commands - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

3
Commands by Subsystem

:TRIGger:GLITch Commands

Table 68 :TRIGger:GLITch Commands Summary
Command
:TRIGger:GLITch:GREater
than <greater than
time>[suffix] (see
page
384)
:TRIGger:GLITch:LESStha
n <less than
time>[suffix] (see
page
385)
:TRIGger:GLITch:LEVel
<level> [<source>] (see
page
386)
:TRIGger:GLITch:POLarit
y <polarity> (see
page
387)
:TRIGger:GLITch:QUALifi
er <qualifier> (see
page
388)
382
Query
:TRIGger:GLITch:GREater
than? (see
page
384)
:TRIGger:GLITch:LESStha
n? (see
page
385)
:TRIGger:GLITch:LEVel?
(see
page
386)
:TRIGger:GLITch:POLarit
y? (see
page
387)
:TRIGger:GLITch:QUALifi
er? (see
page
388)
Options and Query Returns
<greater than time> ::=
floating-point number from 5 ns to
10 seconds in NR3 format
[suffix] ::= {s | ms | us | ns | ps}
<less than time> ::= floating-point
number from 5 ns to 10 seconds in
NR3 format
[suffix] ::= {s | ms | us | ns | ps}
For internal triggers, <level> ::=
.75 x full-scale voltage from center
screen in NR3 format.
For external triggers, <level> ::= 2
volts with probe attenuation at 1:1
in NR3 format.
For digital channels (MSO models),
<level> ::= 6 V.
<source> ::= {CHANnel<n> | EXTernal}
for DSO models
<source> ::= {CHANnel<n> |
DIGital0,..,DIGital15} for MSO
models
<n> ::= 1-2 or 1-4 in NR1 format
<polarity> ::= {POSitive | NEGative}
<qualifier> ::= {GREaterthan |
LESSthan | RANGe}
Agilent 6000 Series Oscilloscopes Programmer's Reference

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents