Ovlenable (Overload Event Enable Register) - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

3
Commands by Subsystem

:OVLenable (Overload Event Enable Register)

Command Syntax
:OVLenable <enable_mask>
<enable_mask> ::= 16-bit integer
The overload enable mask is an integer representing an input as described
in the following table.
The :OVLenable command sets the mask in the Overload Event Enable
Register and enables the reporting of the Overload Event Register. If an
overvoltage is sensed on a 50
to 1 M
Operation Status Register.
You can set analog channel input impedance to 50
N O T E
bandwidth oscilloscope models. On these same bandwidth models, if there are only two
analog channels, you can also set external trigger input impedance to 50
Table 45 Overload Event Enable Register (OVL)
Bit
15-11
10
9
8
126
(see
page
564)
input impedance. If enabled, such an event will set bit 11 in the
Description
---
External Trigger Fault
Channel 4 Fault
Channel 3 Fault
input, the input will automatically switch
on the 300 MHz, 500 MHz, and 1 GHz
When Set (1 = High = True), Enables:
(Not used.)
Event when fault occurs on External Trigger input.
Event when fault occurs on Channel 4 input.
Event when fault occurs on Channel 3 input.
Agilent 6000 Series Oscilloscopes Programmer's Reference
.