:TRIGger:FLEXray:FRAMe:CCRepetition

Command Syntax
:TRIGger:FLEXray:FRAMe:CCRepetition <cycle_count_repetition>
<cycle_count_repetition> ::= {ALL | <rep #>}
<rep #> ::= integer from 2-64
The :TRIGger:FLEXray:FRAMe:CCRepetition command sets the repetition
number of the FlexRay cycle count (in the frame header) to trigger on. The
cycle count repetition setting is only valid when the FlexRay trigger mode
is set to FRAME.
This command is only valid when the FLEXray triggering and serial decode option (Option
N O T E
FRS) has been licensed.
Query Syntax
:TRIGger:FLEXray:FRAMe:CCRepetition?
The :TRIGger:FLEXray:FRAMe:CCRepetition? query returns the current
cycle count repetition setting for the FlexRay frame trigger setup.
Return Format
<cycle_count_repetition><NL>
<cycle_count_repetition> ::= {ALL | <rep #>}
<rep #> ::= integer from 2-64
Errors
See Also
374
(see
page
564)
"- 241, Hardware missing"
"Introduction to :TRIGger Commands"
":TRIGger:FLEXray:TRIGger"
on page 535
on page 332
on page 381