Channel:threshold - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

5
Obsolete and Discontinued Commands

:CHANnel:THReshold

Command Syntax
:CHANnel:THReshold <channel group>, <threshold type> [, <value>]
<channel group> ::= {POD1 | POD2}
<threshold type> ::= {CMOS | ECL | TTL | USERdef}
<value> ::= voltage for USERdef in NR3 format [volt_type]
[volt_type] ::= {V | mV (-3) | uV (-6)}
The :CHANnel:THReshold command sets the threshold for a group of
channels. The threshold is either set to a predefined value or to a
user- defined value. For the predefined value, the voltage parameter is
ignored.
The :CHANnel:THReshold command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :POD<n>:THReshold command (see
:DIGital<n>:THReshold command (see
Query Syntax
:CHANnel:THReshold? <channel group>
The :CHANnel:THReshold? query returns the voltage and threshold text for
a specific group of channels.
Return Format
<threshold type> [, <value>]<NL>
<threshold type> ::= {CMOS | ECL | TTL | USERdef}
<value> ::= voltage for USERdef (float 32 NR3)
• CMOS = 2.5V
N O T E
• TTL = 1.5V
• ECL = - 1.3V
• USERdef ::= - 6.0V to 6.0V
498
(see
page
564)
page
192) for the 6000 Series oscilloscopes.
Agilent 6000 Series Oscilloscopes Programmer's Reference
page
290) or

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents