:WAVeform:PREamble

Query Syntax
:WAVeform:PREamble?
The :WAVeform:PREamble query requests the preamble information for the
selected waveform source. The preamble data contains information
concerning the vertical and horizontal scaling of the data of the
corresponding channel.
Return Format
<preamble_block><NL>
<preamble_block> ::= <format 16-bit NR1>,
<format> ::= 0 for BYTE format, 1 for WORD format, 2 for ASCii format;
<type> ::= 2 for AVERage type, 0 for NORMal type, 1 for PEAK detect
<count> ::= Average count or 1 if PEAK or NORMal; an integer in NR1
Agilent 6000 Series Oscilloscopes Programmer's Reference
(see
page
564)
<type 16-bit NR1>,
<points 32-bit NR1>,
<count 32-bit NR1>,
<xincrement 64-bit floating point NR3>,
<xorigin 64-bit floating point NR3>,
<xreference 32-bit NR1>,
<yincrement 32-bit floating point NR3>,
<yorigin 32-bit floating point NR3>,
<yreference 32-bit NR1>
an integer in NR1 format (format set by :WAVeform:FORMat).
type; an integer in NR1 format (type set by :ACQuire:TYPE).