Obsolete and Discontinued Commands

:MEASure:TMIN

Command Syntax
:MEASure:TMIN [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:TMIN command installs a screen measurement and starts an
X- at- Min- Y measurement on the selected waveform. If the optional source
is specified, the current source is modified.
The :MEASure:TMIN command is obsolete and is provided for backward compatibility to
N O T E
previous oscilloscopes. Use the :MEASure:XMIN command (see
Query Syntax
:MEASure:TMIN? [<source>]
The :MEASure:TMIN? query returns the horizontal axis value at which the
minimum vertical value occurs on the current source. If the optional
source is specified, the current source is modified. If all channels are off,
the query returns 9.9E+37.
Return Format
<value><NL>
<value> ::= time at minimum in NR3 format
See Also
516
(see
page
564)
"Introduction to :MEASure Commands"
":MEASure:TMAX"
on page 515
":MEASure:XMAX"
on page 285
":MEASure:XMIN"
on page 286
on page 248
Agilent 6000 Series Oscilloscopes Programmer's Reference
page
286) instead.