:TRIGger:CAN:SIGNal:BAUDrate

Command Syntax
:TRIGger:CAN:SIGNal:BAUDrate <baudrate>
<baudrate> ::= integer in NR1 format
<baudrate> ::= {10000 | 20000 | 33300 | 50000 | 62500 | 83300 | 100000
The :TRIGger:CAN:SIGNal:BAUDrate command sets the standard baud rate
of the CAN signal from 10 kb/s to 1 Mb/s. If a non- standard baud rate is
sent, the baud rate will be set to the next highest standard rate.
If the baud rate you select does not match the system baud rate, false
triggers may occur.
Query Syntax
:TRIGger:CAN:SIGNal:BAUDrate?
The :TRIGger:CAN:SIGNal:BAUDrate? query returns the current CAN baud
rate setting.
Return Format
<baudrate><NL>
<baudrate> ::= integer = {10000 | 20000 | 33300 | 50000 | 62500
See Also
350
(see
page
564)
| 125000 | 250000 | 500000 | 800000 |1000000}
"Introduction to :TRIGger Commands"
":TRIGger:MODE"
on page 338
":TRIGger:CAN:TRIGger"
":TRIGger:CAN:SIGNal:DEFinition"
":TRIGger:CAN:SOURce"
| 83300 | 100000 | 125000 | 250000 | 500000
| 800000 |1000000}
on page 332
on page 352
on page 529
on page 351
Agilent 6000 Series Oscilloscopes Programmer's Reference