:WAVeform:POINts:MODE

Command Syntax
:WAVeform:POINts:MODE <points_mode>
<points_mode> ::= {NORMal | MAXimum | RAW}
The :WAVeform:POINts:MODE command sets the data record to be
transferred with the :WAVeform:DATA? query.
For the analog or digital sources, there are two different records that can
be transferred:
• The first is the raw acquisition record. The maximum number of points
• The second is referred to as the measurement record and is a 1000
If the <points_mode> is NORMal, the measurement record is retrieved.
If the <points_mode> is RAW, the raw acquisition record is used. Under
some conditions, such as when the oscilloscope is running, this data
record is unavailable.
If the <points_mode> is MAXimum, whichever record contains the
maximum amount of points is used. Usually, this is the raw acquisition
record. But, if the raw acquisition record is unavailable (for example,
when the oscilloscope is running), or if the reconstruction filter (Sin(x)/x
interpolation) is in use, the measurement record may have more data. If
data is being retrieved as the oscilloscope is stopped and as the data
displayed is changing, the data being retrieved can switch between the
measurement and raw acquisition records.
Considerations
• The instrument must be stopped (see the :STOP command (see
for MAXimum or
RAW data
retrieval
• :TIMebase:MODE must be set to MAIN.
• :ACQuire:TYPE must be set to NORMal, AVERage, or HRESolution. If
• MAXimum or RAW will allow up to 8,000,000 points to be returned. The
Agilent 6000 Series Oscilloscopes Programmer's Reference
(see
page
564)
available in this record is returned by the :ACQuire:POINts? query. The
raw acquisition record can only be transferred when the oscilloscope is
not running and can only be retrieved from the analog or digital
sources.
point (maximum) representation of the raw acquisition record. The
measurement record can be retrieved at any time, from any source.
page
135) or the :DIGitize command (see
subsystem) in order to return more than 1000 points.
AVERage, :ACQuire:COUNt must be set to 1 in order to return more
than 1000 points.
number of points returned will vary as the instrument's configuration is
changed. Use the :WAVeform:POINts? MAXimum query to determine the
maximum number of points that can be retrieved at the current
settings.