Measure:vamplitude - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

3
Commands by Subsystem

:MEASure:VAMPlitude

Command Syntax
:MEASure:VAMPlitude [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:VAMPlitude command installs a screen measurement and
starts a vertical amplitude measurement. If the optional source parameter
is specified, the current source is modified.
Query Syntax
:MEASure:VAMPlitude? [<source>]
The :MEASure:VAMPlitude? query measures and returns the vertical
amplitude of the waveform. To determine the amplitude, the instrument
measures Vtop and Vbase, then calculates the amplitude as follows:
Return Format
<value><NL>
<value> ::= the amplitude of the selected waveform in NR3 format
See Also
276
(see
page
564)
vertical amplitude = Vtop - Vbase
"Introduction to :MEASure Commands"
":MEASure:SOURce"
":MEASure:VBASe"
on page 278
":MEASure:VTOP"
on page 284
":MEASure:VPP"
on page 281
on page 248
on page 270