3
Commands by Subsystem
:TRIGger:SWEep
Command Syntax
:TRIGger:SWEep <sweep>
<sweep> ::= {AUTO | NORMal}
The :TRIGger:SWEep command selects the trigger sweep mode.
When AUTO sweep mode is selected, a baseline is displayed in the absence
of a signal. If a signal is present but the oscilloscope is not triggered, the
unsynchronized signal is displayed instead of a baseline.
When NORMal sweep mode is selected and no trigger is present, the
instrument does not sweep, and the data acquired on the previous trigger
remains on the screen.
This feature is called "Mode" on the instrument's front panel.
N O T E
Query Syntax
:TRIGger:SWEep?
The :TRIGger:SWEep? query returns the current trigger sweep mode.
Return Format
<sweep><NL>
<sweep> ::= current trigger sweep mode
See Also
•
342
(see
page
564)
"Introduction to :TRIGger Commands"
on page 332
Agilent 6000 Series Oscilloscopes Programmer's Reference