External:range - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

3
Commands by Subsystem

:EXTernal:RANGe

Command Syntax
:EXTernal:RANGe <range>[<suffix>]
<range> ::= vertical full-scale range value in NR3 format
<suffix> ::= {V | mV}
The :EXTernal:RANGe command is provided for product compatibility. The
range can only be set to 5.0 V when using 1:1 probe attenuation. If the
probe attenuation is changed, the range value is multiplied by the probe
attenuation factor.
Query Syntax
:EXTernal:RANGe?
The :EXTernal:RANGe? query returns the current full- scale range setting
for the external trigger.
Return Format
<range_argument><NL>
<range_argument> ::= external trigger range value in NR3 format = (5.0 V
) * (probe attenuation factor)
See Also
210
(see
page
564)
"Introduction to :EXTernal Trigger Commands"
":EXTernal:PROBe"
on page 206
"Introduction to :TRIGger Commands"
on page 332
Agilent 6000 Series Oscilloscopes Programmer's Reference
on page 203