Trigger[:Edge]:Reject - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

:TRIGger[:EDGE]:REJect

Command Syntax
:TRIGger[:EDGE]:REJect <reject>
<reject> ::= {OFF | LFReject | HFReject}
The :TRIGger[:EDGE]:REJect command turns the low- frequency or
high- frequency reject filter on or off. You can turn on one of these filters
at a time.
• The high frequency reject filter adds a 50 kHz low- pass filter in the
• The low frequency reject filter adds a 50 kHz high- pass filter in series
The :TRIGger[:EDGE]:REJect and the :TRIGger[:EDGE]:COUPling selections are coupled.
N O T E
Changing the setting of the :TRIGger[:EDGE]:COUPling can change the COUPling setting.
Query Syntax
:TRIGger[:EDGE]:REJect?
The :TRIGger[:EDGE]:REJect? query returns the current status of the reject
filter.
Return Format
<reject><NL>
<reject> ::= {OFF | LFR | HFR}
See Also
Agilent 6000 Series Oscilloscopes Programmer's Reference
(see
page
564)
trigger path to remove high frequency components from the trigger
waveform. Use the high frequency reject filter to remove high- frequency
noise, such as AM or FM broadcast stations, from the trigger path.
with the trigger waveform to remove any unwanted low frequency
components from a trigger waveform, such as power line frequencies,
that can interfere with proper triggering.
"Introduction to :TRIGger Commands"
":TRIGger:HFReject"
on page 336
":TRIGger[:EDGE]:COUPling"
Commands by Subsystem
on page 332
on page 365