:MARKer:XDELta

Query Syntax
:MARKer:XDELta?
The MARKer:XDELta? query returns the value difference between the
current X1 and X2 cursor positions.
Xdelta = (Value at X2 cursor) - (Value at X1 cursor)
If the front-panel cursors are off or are set to Binary or Hex Mode, the marker position
N O T E
values are not defined. Make sure to set :MARKer:MODE to MANual to put the cursors in
the front-panel Normal mode.
Return Format
<value><NL>
<value> ::= difference value in NR3 format.