Channel<N>:Scale - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

:CHANnel<n>:SCALe
Command Syntax
:CHANnel<n>:SCALe <scale>[<suffix>]
<scale> ::= vertical units per division in NR3 format
<suffix> ::= {V | mV}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:SCALe command sets the vertical scale, or units per
division, of the selected channel. When using 1:1 probe attenuation, legal
values for the scale range from:
• 1 mV to 5 V for the 100 MHz models.
• 2 mV to 1 V for the 300 MHz – 1 GHz models with the input impedance
If the probe attenuation is changed, the scale value is multiplied by the
probe's attenuation factor.
Query Syntax
:CHANnel<n>:SCALe?
The :CHANnel<n>:SCALe? query returns the current scale setting for the
specified channel.
Return Format
<scale value><NL>
<scale value> ::= vertical units per division in NR3 format
See Also
Agilent 6000 Series Oscilloscopes Programmer's Reference
(see
page
564)
set to 50
.
"Introduction to :CHANnel<n> Commands"
":CHANnel<n>:RANGe"
":CHANnel<n>:PROBe"
on page 182
on page 177
Commands by Subsystem
on page 168