3
Commands by Subsystem

:MEASure:SDEViation

Command Syntax
:MEASure:SDEViation [<source>]
<source> ::= {CHANnel<n> | FUNCtion | MATH}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :MEASure:SDEViation command installs a screen measurement and
starts std deviation measurement. If the optional source parameter is
specified, the current source is modified.
This command is not available if the source is FFT (Fast Fourier Transform).
N O T E
Query Syntax
:MEASure:SDEViation? [<source>]
The :MEASure:SDEViation? query measures and outputs the std deviation
of the selected waveform. The oscilloscope computes the std deviation on
all displayed data points.
Return Format
<value><NL>
<value> ::= calculated std deviation value in NR3 format
See Also
268
(see
page
564)
"Introduction to :MEASure Commands"
":MEASure:SOURce"
on page 248
on page 270