Channel<N>:Bwlimit - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

3
Commands by Subsystem
:CHANnel<n>:BWLimit
Command Syntax
:CHANnel<n>:BWLimit <bwlimit>
<bwlimit> ::= {{1 | ON} | {0 | OFF}
<n> ::= {1 | 2 | 3 | 4} for the four channel oscilloscope models
<n> ::= {1 | 2} for the two channel oscilloscope models
The :CHANnel<n>:BWLimit command controls an internal low- pass filter.
When the filter is on, the bandwidth of the specified channel is limited to
approximately 25 MHz.
Query Syntax
:CHANnel<n>:BWLimit?
The :CHANnel<n>:BWLimit? query returns the current setting of the
low- pass filter.
Return Format
<bwlimit><NL>
<bwlimit> ::= {1 | 0}
See Also
170
(see
page
564)
"Introduction to :CHANnel<n> Commands"
on page 168
Agilent 6000 Series Oscilloscopes Programmer's Reference