Acquire:type - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

3
Commands by Subsystem

:ACQuire:TYPE

Command Syntax
:ACQuire:TYPE <type>
<type> ::= {NORMal | AVERage | HRESolution | PEAK}
The :ACQuire:TYPE command selects the type of data acquisition that is to
take place. The acquisition types are: NORMal, AVERage, HRESolution, and
PEAK.
• The :ACQuire:TYPE NORMal command sets the oscilloscope in the
• The :ACQuire:TYPE AVERage command sets the oscilloscope in the
• The :ACQuire:TYPE HRESolution command sets the oscilloscope in the
• The :ACQuire:TYPE PEAK command sets the oscilloscope in the peak
The obsolete command ACQuire:TYPE:REALtime is functionally equivalent to sending
N O T E
ACQuire:MODE RTIME; TYPE NORMal.
Query Syntax
:ACQuire:TYPE?
The :ACQuire:TYPE? query returns the current acquisition type.
Return Format
<acq_type><NL>
<acq_type> ::= {NORM | AVER | HRES | PEAK}
See Also
148
(see
page
564)
normal mode.
averaging mode. You can set the count by sending the :ACQuire:COUNt
command followed by the number of averages. In this mode, the value
for averages is an integer from 1 (smoothing) to 65536. The COUNt
value determines the number of averages that must be acquired.
high- resolution mode (also known as smoothing). This mode is used to
reduce noise at slower sweep speeds where the digitizer samples faster
than needed to fill memory for the displayed time range.
For example, if the digitizer samples at 200 MSa/s, but the effective
sample rate is 1 MSa/s (because of a slower sweep speed), only 1 out of
every 200 samples needs to be stored. Instead of storing one sample
(and throwing others away), the 200 samples are averaged together to
provide the value for one display point. The slower the sweep speed,
the greater the number of samples that are averaged together for each
display point.
This command is functionally equivalent to :ACQuire:TYPE AVERage
and :ACQuire:COUNt 1.
detect mode. In this mode, :ACQuire:COUNt has no meaning.
"Introduction to :ACQuire Commands"
on page 138
Agilent 6000 Series Oscilloscopes Programmer's Reference

Hide quick links:

Advertisement

Table of Contents
loading