Measure:tstop - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

5
Obsolete and Discontinued Commands

:MEASure:TSTOp

Command Syntax
:MEASure:TSTOp <value> [suffix]
<value> ::= time at the stop marker in seconds
[suffix] ::= {s | ms | us | ns | ps}
The :MEASure:TSTOp command moves the stop marker (X2 cursor) to the
specified time with respect to the trigger time.
The short form of this command, TSTO, does not follow the defined Long Form to Short
N O T E
Form Truncation Rules (see
both TSTArt and TSTOp, so sending TST for the TSTOp command produces an error.
The :MEASure:TSTOp command is an obsolete command provided for compatibility to
N O T E
previous oscilloscopes. Use the :MARKer:X2Position command (see
Query Syntax
:MEASure:TSTOp?
The :MEASure:TSTOp? query returns the time at the stop marker (X2
cursor).
Return Format
<value><NL>
<value> ::= time at the stop marker in NR3 format
See Also
518
(see
page
564)
"Introduction to :MARKer Commands"
"Introduction to :MEASure Commands"
":MARKer:X1Position"
":MARKer:X2Position"
":MARKer:XDELta"
on page 239
":MEASure:TDELta"
on page 513
":MEASure:TSTArt"
on page 517
page
566). The normal short form "TST" would be the same for
on page 233
on page 248
on page 235
on page 237
Agilent 6000 Series Oscilloscopes Programmer's Reference
page
237) instead.