:TIMebase:POSition

Command Syntax
:TIMebase:POSition <pos>
<pos> ::= time in seconds from the trigger to the display reference
The :TIMebase:POSition command sets the time interval between the
trigger event and the display reference point on the screen. The display
reference point is either left, right, or center and is set with the
:TIMebase:REFerence command. The maximum position value depends on
the time/division settings.
This command is an alias for the :TIMebase:DELay command.
N O T E
Query Syntax
:TIMebase:POSition?
The :TIMebase:POSition? query returns the current time from the trigger to
the display reference in seconds.
Return Format
<pos><NL>
<pos> ::= time in seconds from the trigger to the display reference
See Also
Agilent 6000 Series Oscilloscopes Programmer's Reference
(see
page
564)
in NR3 format
in NR3 format
"Introduction to :TIMebase Commands"
":TIMebase:REFerence"
":TIMebase:RANGe"
on page 324
":TIMebase:SCALe"
on page 327
":TIMebase:WINDow:POSition"
":TIMebase:DELay"
on page 527
on page 320
on page 326
on page 329
Commands by Subsystem
3
323