Calibrate Commands - Agilent Technologies MSO6102A Programmer's Reference Manual

6000 series
Hide thumbs Also See for MSO6102A:
Table of Contents

Advertisement

:CALibrate Commands

Utility commands for viewing calibration status and for starting the user
calibration procedure. See
page 159.
Table 49 :CALibrate Commands Summary
Command
n/a
:CALibrate:LABel
<string> (see
page
:CALibrate:STARt (see
page
162)
n/a
n/a
n/a
n/a
Introduction to
The CALibrate subsystem provides utility commands for:
:CALibrate
• Determining the state of the calibration factor protection switch
Commands
• Saving and querying the calibration label string.
• Reporting the calibration time and date.
• Reporting changes in the temperature since the last calibration.
• Starting the user calibration procedure.
Agilent 6000 Series Oscilloscopes Programmer's Reference
Query
:CALibrate:DATE? (see
page
160)
:CALibrate:LABel? (see
161)
page
161)
n/a
:CALibrate:STATus? (see
page
163)
:CALibrate:SWITch? (see
page
164)
:CALibrate:TEMPerature?
(see
page
165)
:CALibrate:TIME? (see
page
166)
(CAL PROTECT).
"Introduction to :CALibrate Commands"
Options and Query Returns
<return value> ::=
<day>,<month>,<year>; all in NR1
format
<string> ::= quoted ASCII string up
to 32 characters
n/a
<return value> ::=
ALL,<status_code>,<status_string>
<status_code> ::= an integer status
code
<status_string> ::= an ASCII status
string
{PROTected | UNPRotected}
<return value> ::= degrees C delta
since last cal in NR3 format
<return value> ::=
<hours>,<minutes>,<seconds>; all in
NR1 format
Commands by Subsystem
on
3
159