:CHANnel<n>:PROTection
Command Syntax
:CHANnel<n>:PROTection[:CLEar]
<n> ::= {1 | 2 | 3 | 4}
When the analog channel input impedance is set to 50
500 MHz, and 1 GHz bandwidth oscilloscope models), the input channels
are protected against overvoltage. When an overvoltage condition is sensed,
the input impedance for the channel is automatically changed to 1 M
The :CHANnel<n>:PROTection[:CLEar] command is used to clear (reset) the
overload protection. It allows the channel to be used again in 50
after the signal that caused the overload has been removed from the
channel input. Reset the analog channel input impedance to 50
":CHANnel<n>:IMPedance"
protection.
Query Syntax
:CHANnel<n>:PROTection?
The :CHANnel<n>:PROTection query returns the state of the input
protection for CHANnel<n>. If a channel input has experienced an
overload, TRIP (tripped) will be returned; otherwise NORM (normal) is
returned.
Return Format
{NORM | TRIP}<NL>
See Also
Agilent 6000 Series Oscilloscopes Programmer's Reference
(see
page
564)
"Introduction to :CHANnel<n> Commands"
":CHANnel<n>:COUPling"
":CHANnel<n>:IMPedance"
":CHANnel<n>:PROBe"
on page 173) after clearing the overvoltage
on page 171
on page 173
on page 177