How to emove the Front Panel Assembly
1. Disconnect the power cord.
2. Remove the front handles (trim strip rst). See Figure 3-1.
3. Pull the center top of the front panel up slightly and pull the center bottom down slightly to
release the two centered catches.
4. Pull the front panel away from the cabinet several inches.
5. Disconnect the ribbon cable from the circuit board.
a. If you are planning to remove the A3, A4, or A5 assemblies later, remove the handle nut
plate on the right side.
b. If you are planning to remove the A6 or A7 assembly, you should remove the handle nut
plate on the left side.
c. If you are planning to remove the A2 assembly, you should remove both handle nut
plates.
6. o remove the RPG knob, pull it o its splined shaft.
7. o remove the RPG, disconnect the ve-wire cable from A1J2. Remove the knob, hex nut,
and washer.
Note
If you are replacing a probe power fuse, they are located in sockets on the A1
front panel assembly. See able 7-3, Item 7.
3-4
Assembly Replacement