To Remove And Replace The Handle; Feet And Tilt Stand - Agilent Technologies 1670G Series Service Manual

Logic analyzers
Hide thumbs Also See for 1670G Series:
Table of Contents

Advertisement

To remove and replace the handle

Remove the two screws in the endcaps, then lift off the handle.
Replace the handle and tighten the screws to 2.4 Nm (21 in-lb.).
To remove and replace the feet and tilt stand
Remove the screws connecting the four rear feet to the instrument.
1
Separate the rear feet from the instrument to remove them.
2
Press the locking tab on the bottom feet, then remove them.
3
Remove the tilt stand from the bottom front feet by sliding each foot off of the stand.
4
Reverse this procedure to install the feet and tilt stand.
5
To remove and replace the cover
Turn the power off and remove the power cord.
1
Remove the probe plate and disconnect the logic analyzer cables from the rear panel.
2
Using the previous procedures, remove the handle and the four rear feet.
3
Remove the seven screws from the front molding, then slide the molding forward to
4
remove it.
Remove the cover.
5
To remove the cover, set the instrument upright and facing toward you. Slide the chassis
toward the front, out of the cover, and set it on a static-safe work area.
Reverse this procedure to install the cover.
6
Check that all assemblies are properly installed before installing the cover.
When installing the chassis in the cover, check that the tabs located at the bottom, rear of the
cover align with the holes in the rear panel.
Replacing Assemblies
To remove and replace the handle
6–5